· Bảng điều khiển cảm ứng TFT-LCD 5,7 inch 640 x 480
– Đầu vào đa năng (Cặp nhiệt điện, RTD, VDC)
– Chế độ xem xu hướng có sẵn
– Bộ nhớ trong 80M byte và hỗ trợ thẻ nhớ SD/SDHC
– Giao tiếp RS 485/422 ( Modbus, PC-link )
· Suffix code
| Model | Code | Information | ||
| GR200 – | Graphic recorder 145(W) x 145(H) x 173.5(D) ㎜ | |||
| Input channel | 2 | 2 channel | ||
| 4 | 4 channel | |||
| 8 | 8 channel | |||
| 12 | 12 channel | |||
| External contact input & output (DI / DO) |
N | None | ||
| 1 | DI 2 contact + DO 6 contacts | |||
| 2 | DI 4 contact + DO 12 contacts | |||
| Communication function | 0 | RS422/485 | ||
· Specification
| [ Range configuration for the input types ] |
| Input type | Measurement range(°C) | Measurement range(°F) | Degree | ||
| Thermoresistor (RTD) | Pt100Ω | Pt-0 | -200 ~ 640 | -300 ~ 1180 | ±0.15 % of F.S ±1digit |
| Pt-1 | -1 -200.0 ~ 640.0 | -300.0 ~ 1180.0 | |||
| Pt-2 | -200.00 ~ 640.00 | -300.0 ~ 1180.0 | |||
| KPt100Ω | KPt-0 | -200 ~ 500 | -300 ~ 1000 | ||
| KPt-1 | -200.0 ~ 500.0 | -300.0 ~ 1000.0 | |||
| KPt-2 | -100.00 ~ 150.00 | -300.0 ~ 1000.0 | |||
| Thermocouple (TC) | K | K-0 | -200 ~ 1370 | -300 ~ 2500 | ±0.15 % of F.S ±1digit |
| K-1 | -200.0 ~ 1370.0 | -300 ~ 2500 | ±0.15 % of F.S ±1digit(*2) | ||
| J | -200.0 ~ 1200.0 | -300 ~ 2300 | ±0.15 % of F.S ±1digit(*2) | ||
| E | -200.0 ~ 1000.0 | -300 ~ 1800 | ±0.15 % of F.S ±1digit(*2) | ||
| T | -200.0 ~ 400.0 | -300 ~ 750 | ±0.15 % of F.S ±1digit(*3) | ||
| R | 0.0 ~ 1700.0 | 0 ~ 3100 | ±0.15 % of F.S ±1digit(*4) | ||
| B | 0.0 ~ 1800.0 | 0 ~ 3300 | ±0.15 % of F.S ±1digit (*1) | ||
| S | 0.0 – 1700.0 | 0 ~ 3100 | ±0.15 % of F.S ±1digit(*4) | ||
| L | -200.0 ~ 900.0 | -300 ~ 1300 | ±0.15 % of F.S ±1digit(*2) | ||
| N | -200.0 ~ 1300.0 | -300 ~ 2400 | ±0.15 % of F.S ±1digit(*3) | ||
| U | -200.0 ~ 400.0 | -300 ~ 750 | ±0.15 % of F.S ±1digit(*3) | ||
| Wre 5-26 | 0.0 ~ 2300.0 | 0 ~ 4200 | ±0.15 % of F.S ±1digit(*4) | ||
| PL-Ⅱ | 0.0 ~ 1390.0 | 0 ~ 2500 | ±0.15 % of F.S ±1digit | ||
| VDC | -10 – 20 ㎷ | -10.00 ~ 20.00 | ±0.15 % of F.S ±1digit * The current input (4~20㎃ DC) is available when you use resistance 250Ω (0.5W/0.1%) on input terminals |
||
| 0 – 100 ㎷ | 0.00 ~ 100.00 | ||||
| 1 – 5 V | 1.00 ~ 5.00 | ||||
| 0 – 30 V | 0.00 ~ 30.00 | ||||
| (*1) 0 ~ 400℃ : ±5% of F.S ± 1digit | (*3) -200 ~ -150℃ : ±0.4% of F.S ± 1digit |
| (*2) -200 ~ -150℃ : ±0.2% of F.S ± 1digit | -150 ~ -100℃ : ±0.2% of F.S ± 1digit |
| (*4) 0 ~ 200℃ : ±0.2% of F.S ± 1digit |
| [ Specification for the input ] |
| VDC | Thermocouple (TC) | K, J, E, T, R, S, B, N | IEC 584 |
| PL-Ⅱ, Wre 5-26 | ASTM E988 | ||
| U, L | DIN 43710, IEC 751 | ||
| Thermoresistor (RTD) | Pt100 Ω | DIN IEC 751 | |
| KPt100 Ω | JIS C1604-1989 (OLD) |
| [Power input] |
| Power voltage | 100 – 240 V AC Voltage fluctuation rate ±10 % |
| Power frequency | 50 – 60 ㎐ |
| Power consumption | 22 VA max |
| Maximum internal fuse ratings | 250 V AC |
| Internal voltage | Primary terminal* and secondary terminal** : Minimum 1500 VAC for 1 minute Primary terminal* and FG terminal : Minimum 1500 VAC for 1 minute Secondary terminal** and FG terminal : Minimum 1500 VAC for 1 minute Secondary terminal** and secondary terminal** : Minimum 500 VAC for 1 minute * Primary terminal : Power terminal (except the FG terminal) and the relay output terminal ** Secondary terminal : Sensor input terminal, contact input terminal, communication terminal |
| Insulation resistor | 20 ㏁ between the power terminal and the FG terminal or 500 VDC |
| [Sensor input] |
| Number of channels | 2, 4, 8, 12 (Refer to the type configuration) |
| Input type | 2 thermoresistors (Pt-100, KPt-100) 12 thermocouples (K, J, E, T, R, B, S, L, N, U, Wre 5-26, PL-II) 4 VDC (-10 – 20 mV, 0 – 100 mV, 1 – 5V, 0 – 30 V) |
| Sampling period | 250 ㎳ |
| Current to measure the thermoresistor (RTD) | About 0.21 ㎃ |
| Input resistor | Thermocouple : More than 1 ㏁, VDC : More than 1 ㏁ |
| Allowable wiring resistor | Thermoresistor : Maximum 100Ω/wire (The RTD is up to 10Ω/wire for the range of -100.00 – .150.00) Thermocouple : Less than 200 Ω VDC : Less than 2 ㏀ |
| Impact of the wiring resistor | Thermoresistor : ±0.3 °C/10 Ω (The 3 lines have the same wiring resistors) |
| Allowable input voltage | Thermocouple : Less than ±10 V DC, VDC : Less than ±33 V DC |
| Degree | ±0.15 % of F.S, ±1 digit (Except the RJC temperature error) ※ Refer to the input table |
| The error in the reference junction compensation (RJC) | ±2.0 ℃ (0 ~ 50 ℃) |
| Sensor short detection (Burn-out) | UP-Scale for the short |
| [Contact input] |
| Maximum input | 4 |
| Input method | No voltage contact input |
| On/Off detection resistor | Consider on less than 1 ㏀ minimum and off larger than 10 ㏀ |
| Minimum detection time | 0.25 second |
| [Alarm output] |
| Maximum number of outputs | 12 |
| Output type | Relay output |
| Maximum ratings | 5 A 250 V AC, 5 A 30 V DC |
| Recommended ratings | 3 A 250 V AC, 3 A 30 V DC |
| Relay life | 50,000 times at the maximum ratings, 100,000 times at the recommended ratings |
| [Display specification] |
| Display | TFT color LCD (115.2 × 86.4 ㎜, resolution : 640 × 480 pixel, LED Backlight) |
| Backlight life | 40,000 hours |
| Language in use | Korean, English, Chinese (simplified) |
| [Memory specification] |
| Save function | Internal memory – volatile memory : 900 KB – Save up to 4 hours with the interval of 1 second – Non-volatile memory : 80 MB . Save up to 15 days with the interval of 1 second External memory – SD card (8GB) : Save up to a 4 years with the interval of 1 second ※ Support the SDHC |
| Save period | User configuration (1, 2, 5, 10, 20, 30, 60, 120 seconds) |
| Memory information | Measurement from each channel, Burn-out, DI (contact input), ALARM, relay output status |
| [Use environment] |
| Ambient temperature | 0 ~ 50 °C |
| Temperature change | 10 °C/h or below |
| Ambient humidity | 20 ~ 90 % RH (no condensation) |
| [Storage environment] |
| Ambient temperature | -20 ~ 70 °C |
| Temperature change | Less than 20 °C/h |
| Ambient humidity | 5 – 95 % RH (no condensation) |
| [Impact from the ambient temperature] |
| VDC, thermocouple sensor | ±0.01 % of F.S / °C |
| Thermoresistor sensor | ±0.06 °C/°C |